GD3128 Series Insulation Resistance Tester
GD3128 seri Insulation Resistance Tester nduweni fungsi testing sampurna saka macem-macem paramèter resistance insulasi lan kemampuan anti-gangguan banget, kang bisa digunakake kanggo nyoba resistance jampel saka peralatan listrik voltase dhuwur kapasitas gedhe lan garis transmisi ing lingkungan induktif kuwat kayata gardu.Instrumen nganggo layar tutul warna sing bener, kabeh data tes lan daya baterei ditampilake ing layar sing padha kanthi cepet, lan kombinasi tombol lan tombol nggawe operasi gampang banget.Pangguna ora perlu apal cara operasi.Klik "HELP" ing layar kanggo pop munggah cara operasi, kang bisa èfèktif nuntun pangguna kanggo operate instrument.Instrumen kasebut nduweni fungsi ngeculake cepet, sing bisa ngeculake voltase obyek sing diuji kanthi cepet tanpa sirkuit discharge tambahan.Monitor voltase instrumen kanthi otomatis bisa ngawasi voltase urip obyek sing diuji.Yen voltase ngluwihi 36V, tes kasebut bakal dilarang kanthi otomatis, sing bisa nglindhungi instrumen lan operator kanthi efektif.
Instrumen kasebut nggunakake paket baterei lithium sing bisa diisi ulang kanthi kapasitas gedhe.Timer tes instrumen kanthi otomatis nyathet wektu tes lan nyimpen asil tes kanthi tanggal lan wektu.Layar tutul bisa gampang ngakses data sajarah sing direkam.Antarmuka USB sing terisolasi kanthi aman bisa ngunggah data tes menyang PC kanthi aman.Instrumen kasebut dilengkapi garis tes kanthi insulasi kaping pindho lan lapisan pelindung, lan garis voltase dhuwur dilengkapi klip lan pancing buaya sing bisa diganti, sing bisa adaptasi karo macem-macem papan uji.Instrumen kasebut nganggo struktur cangkang lapisan ganda sing padhet, lan kelas perlindungan saka kothak njaba yaiku IP65, sing bisa nyegah invasi kelembapan lan bledug sajrone transportasi lan panyimpenan, lan kanthi efektif nglindhungi instrumen saka tabrakan.
seri GD3128 wis sawetara saka sudhut, résolusi dhuwur, operasi trep, sturdiness lan kekiatan, akurasi lan linuwih, kinerja stabil, lan kemampuan anti-gangguan kuwat.Instrumen kasebut nduweni uji resistensi insulasi (IR);uji indeks polarisasi (PI);tes rasio penyerapan (DAR);ramp test (RAMP);test filter (FR, 10S, 20S, 30S, 40S papat pilihan);tes voltase (V);test kapasitansi (uF);test saiki (nA) lan fungsi liyane.Instrumen kasebut cocog kanggo tes resistensi insulasi peralatan listrik, saluran transmisi lan peralatan gedhe liyane ing lingkungan kanthi kapasitas gedhe, voltase dhuwur lan induksi kuwat.Iki minangka piranti sing umum digunakake lan dibutuhake kanggo tenaga listrik, telekomunikasi, meteorologi, ruangan komputer, lapangan minyak, instalasi lan pangopènan elektromekanik, lan departemen sumber daya ing perusahaan industri lan pertambangan.
●Arus hubung singkat maksimum yaiku 7mA.
●Insulation resistance test (IR), polarization index test (PI), absorption ratio test (DAR).
●Ramp test mode (RAMP), filter test mode (FR), tes kapasitansi (uF)
●Fungsi monitor voltase, kanthi otomatis ngawasi voltase daya obyek sing diuji.
●Fungsi wektu tes, kanthi otomatis ngrekam wektu tes.
●Fungsi discharge otomatis, kanthi otomatis lan cepet ngeculake obyek tes sawise tes, tanpa sirkuit discharge eksternal.
●Fungsi mateni otomatis.
●Layar warna tutul gedhe.
●Struktur cangkang lapisan ganda, kuat lan tahan lama, kelas perlindungan IP65.
●Paket baterei lithium sing bisa diisi ulang kanthi kapasitas dhuwur.
●Fungsi panyimpenan, kanthi otomatis bisa nyimpen 1000 klompok data tes wektu nyata kanthi tanggal lan wektu tes lan wektu tes.
●Fungsi upload, dilengkapi antarmuka USB, data rekaman bisa diunggah menyang komputer kanggo analisis statistik data.
Model produk | Tegangan Rated | Range Resistance Insulation | Short Circuit Saiki |
GD3128A | 250V,500V,1kV,2.5kV,5kV | 0,50MΩ~10,0TΩ | ≥7mA |
GD3128B | 250V,500V,1kV,2.5kV,5kV,10 kV | 0.50MΩ~20.0TΩ | ≥7mA |
GD3128C | 250V,500V,1kV,2.5kV,5kV, 10 kV, 15 kV | 0.50MΩ~30.0TΩ | ≥7mA |
Insulation Resistance Range lan Akurasi
Tegangan Rated | Insulation Resistance Rage | Akurasi | Tetembungane |
250V | 0,50MΩ~50.0GΩ | ±5%rdg±3dgt |
|
50.0GΩ~500GΩ | ±20%rdg±3dgt |
| |
500V | 1.00MΩ~100GΩ | ±5%rdg±3dgt |
|
100GΩ~1.00TΩ | ±20%rdg±3dgt |
| |
1000V | 1.00MΩ~200GΩ | ±5%rdg±3dgt |
|
200GΩ~2.00TΩ | ±20%rdg±3dgt |
| |
2500V | 1.00MΩ~500GΩ | ±5%rdg±3dgt |
|
500GΩ~3.00TΩ | ±20%rdg±3dgt |
| |
5000V | 1.00MΩ~1.00TΩ | ±5%rdg±3dgt |
|
1.00TΩ~10.0TΩ | ±20%rdg±3dgt |
| |
10 KV | 1.00MΩ~2.00TΩ | ±5%rdg±3dgt |
|
2.00TΩ~20.0TΩ | ±20%rdg±3dgt |
| |
15KV | 1.00MΩ~3.00TΩ | ±5%rdg±3dgt |
|
3.00TΩ~30.0TΩ | ±20%rdg±3dgt |
1 TΩ (Tera ohm)=1000GΩ=1012Ω
1 GΩ (Giga ohm)=1000MΩ=109Ω
1 MΩ (Mega ohm)=1000KΩ=106Ω
Fungsi | Pengukuran resistensi isolasi (IR);pengukuran indeks polarisasi (PI);rasio panyerepan (DAR);mode ramp test (RAMP);nyaringfungsi(10S, 20S, 30S, 40S);vpangukuran oltage (V);kapastancepangukuran(uF);pangukuran saiki(nA) |
Power Supply | Baterei lithium sing bisa diisi ulang 6.2Ah |
Tegangan Rated | GD3128A: 250V, 500V, 1kV, 2.5kV, 5kV GD3128B: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV GD3128C: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV, 15kV |
Akurasi Tegangan Output | (5%~10%) ± 10V |
Range Test Resistance Insulation | GD3128: 0,50MΩ~10.0TΩ GD3128B: 0,50MΩ~20.0TΩ GD3128C: 0,50MΩ~30.0TΩ |
Output Short Circuit Saiki | ≥7mA |
Uji Indeks Polarisasi | ya wis |
Tes rasio serapan | ya wis |
Mode Ramp Test | Tes ing langkah 10% saka voltase prasetel nganti voltase prasetel. |
Mode Tes Filter | 10S, 20S, 30S, 40S papat pilihan. |
Tes Tegangan | Range: AC/DC: 0V~1000V; Akurasi: ±5%rdg±3V |
Tes Kapasitansi | Range: 10nF~25uF; Akurasi: ±10%rdg±10nF |
Tes Saiki | Range: 0.01nA~7mA; Akurasi: ±5%rdg±0.5nA |
Monitor Tegangan | Monitor voltase ingtesd obyek, lan ngawasi status discharge sawise test, dilarang kanggo nyoba nalika voltase ngluwihi 36V, nglindhungi instrument lan operator. |
Tes Timer | Wektu tes rekaman otomatis, timingsawetara: 0s~9999s |
Fungsi Panyimpenan | Simpen data tes kanthi otomatis kanthi tanggal lan wektu tes, total 1000 klompok |
Fungsi Upload | Upload data sing disimpen menyang komputer liwat kabel komunikasi USB. |
Daya Baterei Tampilan | Kanthi tampilan daya baterei, nalika voltase baterei kurang bakal ngelingakengisibaterei. |
Mati otomatis | Sawise 15 menit wiwitan bakal mati kanthi otomatis tanpa operasi. |
ukuran | 280mm × 260mm × 160mm |
Bobot | 4900g (kalebu baterei) |
Test Wires | Kawat uji tegangan dhuwur warna abang 1pc (karoa klip buaya lan pancing), kabel uji warna ijo 1pc, kabel test werna ireng 1pc |
Tingkat pangayoman | Tutup casing IP65, bukak casing IP40 |
Lingkungan Kerja | -20 ℃~50 ℃; 80% rh |
Lingkungan Toko | -25 ℃~65 ℃; 80% rh |
Ketahanan Isolasi | 50MΩ(1000V) (antarane sirkuit test lan cangkang) |
Tahan Tegangan | AC 3kV50Hz 1 min(antarane sirkuit test lan cangkang) |
Safety | IEC61010-1, IEC61326-1 |